Espec VAC-100PR Vacuum Oven w/Eyela UT-50 Uni-Trap Moisture Removal System
The unit is guaranteed to work and is in like-new condition with a few nicks and scratches. The oil probably needs to be replaced in the vacuum pump and an overhaul may be necessary. More pictures will be included soon.
The Main Unit's Serial Number Tag Reads:
Model Number: VAC-100PR
Power Requirements: 200 VAC, Single Phase, 50/60 Hz
Max. Amps: 17.2 A
Heater: 2.1 kW
Inside Dimensions: 45 cm W x 45 cm H x 45 cm D
Outside Dimensions: 87 cm W x 148 cm H x 66.2 cm D
Weight: 320 kg.
Date of Manufacture: 2002
Serial Number: 320000980
The Moisture Removal System's Serial Number Tag Reads:
Type: UT-50
Serial Number: 10138623
Power Requirements: 100 VAC, 50/60 Hz, 600 W
The Vacuum Pump's Serial Number Tag Reads:
Type: GLD-200
Pumping Speed:
50 Hz: 200 l/min
60 Hz: 240 l/min
Ultimate Pressure: 6.7 x 10-2 Pa
Oil Capacity: 1100 ml
Serial Number: 75539
Specifications:
Power Supply1: 200 VAC, Single Phase, 2 W, 50/60 Hz
Maximum Power Consumption: 2.75 kVA
Pressure Control System: Fuzzy control
Operating Temperature: +5°C to +35°C (+41°F to +95°F)
Temperature Control Range: +40°C to +200°C (+104°F to +392°F)
Temperature Fluctuations: ±0.5°C (vacuum), ± 1°C (atmospheric)
Temperature Heat-up Rate2: Within 50 min.
Pressure Control Range: 933 to 1 x 102 Pa
Attainment Pressure3: Below 133 Pa
Pressure Pull-down Rate3: Within 7 min. (From atmospheric pressure to 133 Pa)
Pressure Recovery Time: Within 4 min. (Inlet open to atmosphere)
Construction:
Exterior Material: Enameled cold-rolled steel plate
Vacuum Chamber: Stainless steel plate (SUS304)
Interior Material: Stainless steel plate (NSS430M3)
Viewing Window: 324 mm W x 336 mm H (13" x 13.4")
Insulation: Glass wool
Heater: Mica heater
Inlet: R 1/4", max. operating pressure 0.05 MPa (0.5 kg/cm2G)
Outlet: 28 mm external dia. rubber hose connection
Oil Rotary Vacuum Pump:
Motor: 200 VAC, Single Phase, 50/60 Hz, 550 W
Design Exhaust Speed: 200 L/min. (50 Hz), 240 L/min (60 Hz)
Attainment Pressure: 6.7 x 10-2 Pa
Auxiliary Functions: Gas ballast valve, oil mist trap
Fittings: Adjuster feet and casters (free moving, qty. 4)
Effective Inside Capacity: 91 L
Effective Inside Dimensions: 450 mm W x 450 mm H x 450 mm D (18" x 18" x 18")
Outside Dimensions4: 870 mm W x 1450 mm H x 662 mm D (34.8" x 58.0" x 26.5")
Weight: 320 kg. - does not include packaging for shipment
Notes:
1: Voltage fluctuations within ± 10% of rated voltage
2: Time to attain stable temperature at the center of chamber with no specimen, under vacuum with ambient temperature of +23°C and temperature setting of +200°C.
3: Constant temperature inside chamber with no gas generation from specimen.
4: Excluding protrusions.
|