We have a test climate cabinet in excellent condition
Technical data
Brand : Weis Enet
Type : WETK 40.70.1500
Installation identification number : 971160
External connections:
Power supply Voltage : 380/230 volt + ground + zero
Frequention : 50 HZ
Connection Wattages : Max 13 kW
Specific temperature
Temperature range : -50..+ 80 ⁰C
Temperature constant : +/- 2,0⁰C
depend from the time:
Temperature constant : +/- 2,0⁰C
depend from the place:
Heating and cooling rate : Approx 2.0 ⁰ C / min. averaged over range
-40 .. + 70 ⁰ C, with 10 kg. tax
Remaining
Environment temperature : + 10 ⁰C < T < + 30 ⁰C
Refrigerant : R404A / R23, cascade, dual execution
Construction from the closet
Insulation Thickness : 150mm
Internal cladding : Stainless steel 304
External cladding : Painted steel, color RAL 9002
Equipped with front door, left hinges, stainless steel interior, outside painted
steel, color RAL 5015.
Dimensions
Internal (WxDxH) : Circa 110 x 110 x 125 cm
External (WxDxH) : Circa 185 x 230 x 215 cm
Specifications cooling circuits
Installation Identification Number : 971160-WETK
Number of separate cooling circuits : 1
Further indication cooling circuit : Unit 1: M300/M500
: Unit 2: M301/M501
Description of the location
nameplate on the installation : Behind the switch board, against the right -
side of the chamber
Refrigerant type : M300/301 M500/M501
Filling volume (kg) per
cooling circuit : R404A R23
Unit 1 M300/M500 : 5,1 0,85
Unit 2 M301/M501 : 5,3 0,84
Highest pressure (bar) / temperature
(⁰ C) per refrigerant circuit : 22 / 50 13 / -20
Lowest pressure (bar) / temperature
(⁰ C) per refrigerant circuit : 2,4 / -25 2,1 / -60
Maximum power
per compressor (kW) : 2,6 1,5
CFK-detection; Type sensor(S) Not applicable Not applicable
Volume liquid container (kg)
per cooling circuit
Unit 1 M300/M500 : 7,5 Not applicable
Unit 2 M301/M501 : 7,5 Not applicable
Maximum HP-side pressure
(bar) per cooling circuit
Unit 1 M300/M500 : 25 25
Unit 2 M301/M501 : 25 25
Maximum LP-side pressure
(bar) per cooling circuit
Unit 1 M300/M500 : 17,5 17,5
Unit 2 M301/M501 : 17,5 17,5
Air knives of the pressure drain valve
(bar) per cooling circuit Not applicable Not applicable
|